The m360 Select Series climate chamber ICHC was specially developed for precise climate testing under controlled conditions. Even at low temperatures and in humid/CO₂ atmospheres, the ICHC impresses with its constant performance - ideal for pharmaceuticals, the food industry, packaging and building materials testing. A special feature is the compressor technology, which ensures accelerated temperature changes and increased resistance to environmental influences.
The ICHC also supports CO₂-controlled tests to optimally map industry-specific requirements.
Temperature / Working temperature range — from +10°C up to +60°C with humidity
Temperature / Working temperature range — without humidity: 0 to +60 °C
Temperature / Setting temperature range — -10 to +60 °C
Temperature / Setting temperature range — 0 to +60 °C
Temperature / Setting accuracy temperature — 0.1 °C
Temperature / Temperature sensor — 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error
Humidity / Humidity — humidity supply with distilled water from external tank by self-priming pump
Humidity / Humidification — humidification by hot steam generator
Humidity / Humidity — active humidifying and de-humidifying adjustable from 10-80 % rh with digital display of relative humidity - resolution of display 0.1 %, setting accuracy 0.5 %
Control of standard components / Setting range CO 2 — 0 to 20 %CO 2
Control of standard components / Setting accuracy CO2 — 0,1%
Control of standard components / Variation in time CO 2 — +/- 0.2 % CO 2
Control technology / adjustable parameters — temperature (Celsius or Fahrenheit), fan speed, relative humidity, programme time, time zones, summertime/wintertime, CO 2
Control technology / ControlCOCKPIT — TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
Control technology / Language setting — German, English, Spanish, French, Polish, Czech, Hungarian, Italian, Chinese
Control technology / Timer — Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
Control technology / Function SetpointWAIT — the process time does not start until the set temperature is reached
Control technology / Calibration — three freely selectable values each, temperature/humidity, for two freely selectable values CO2
Ventilation / Fan — forced air circulation in working chamber adjustable from 10 to 100 % in 10 % steps
Communication / Documentation — programme stored in case of power failure
Communication / Programming — AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port
Safety / Temperature control — mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 10°C above nominal temperature
Safety / Temperature control — over- and undertemperature monitor TWW, protection class 3.3 or adjustable temperature limiter TWB, protection class 2, selectable on display
Safety / AutoSAFETY — additionally integrated over- and undertemperature protection "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature, compressor in case of undertemperature
Safety / Autodiagnostic system — integral fault diagnostics for temperature and humidity control
Safety / Alarm — visual and acoustic
Heating concept / Air jacket — air jacket heating system for gentle all-around heating
Heating concept / Heating concept — no drying-up of the load caused by the cooling device due to separation of thermal jacket from interior
Heating concept / Cooling — energy-saving use of CFC-free cooling/heating system (refrigerant R290 Propan)
Heating concept / Defrosting — highly efficient automatic defrosting system
Standard equipment / Door — inner glass doors; fully insulated stainless steel door with 2-point locking (compression door lock)
Standard equipment / Standard works calibration certificate — +10°C , +37°C , +30°C / 60% rh; 5%CO 2 at +30°C / 60%rh
Standard equipment / Door — Door-open-recognition
Standard equipment / Standard equipment — CO 2 connection set: hose with coupling and clamp
Standard equipment / Entry port — Entry port (silicone), 40 mm clear diameter, moisture tight, can be closed
by a silicone stopper, standard position at the back
Standard equipment / Internals — 2 stainless steel grid(s), electropolished
Standard equipment / Standard accessories — Water tank including connection hose
Stainless steel interior / Dimensions — w (A) x h (B) x d (C) : 1040 x 1200 x 600 mm (d less 33 mm for fan)
Stainless steel interior / Volume — 749 l
Stainless steel interior / Max. number of internals — 14
Stainless steel interior / Max. loading of chamber — 200 kg
Stainless steel interior / Max. loading per internal — 30 kg
Textured stainless steel casing / Dimensions — w (D) x h (E) x d (F) : 1224 x 1950 x 785 mm (d +56mm door handle)
Textured stainless steel casing / Installation — on lockable castors
Textured stainless steel casing / Housing — rear zinc-plated steel
Electrical data / Voltage
Electrical load — 230 V, 50/60 Hz
approx. 1350 W
Electrical data / Voltage
Electrical load — 115 V, 50/60 Hz
approx. 1350 W