What is hot air sterilisation? Steriliser S makes use of dry sterilisation to ensure laboratory equipment and devices are sterilized. With this method, hot air, devoid of water vapor or moisture, makes sure the device is sterile without needing risk prone disinfection methods.
Temperature / Setting temperature range — +20 to +250 °C
Temperature / Setting accuracy temperature — up to 99.9 °C: 0.1 / from 100 °C: 0.5
Temperature / Working temperature range — at least 10 above ambient temperature to +250°C
Temperature / Temperature sensor — 1 Pt100 sensor DIN class A in 4-wire-circuit
Control technology / ControlCOCKPIT — SingleDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with high-definition TFT-colour display
Control technology / Timer — Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
Control technology / Function SetpointWAIT — the process time does not start until the set temperature is reached
Control technology / Calibration — three freely selectable temperature values
Control technology / adjustable parameters — temperature (Celsius or Fahrenheit), fan speed, air flap position, programme time, time zones, summertime/wintertime
Ventilation / Fan — forced air circulation by 2 quiet air turbines, adjustable in 10 % steps
Ventilation / Fresh air — Admixture of pre-heated fresh air by electronically adjustable air flap
Ventilation / Vent — vent connection with restrictor flap
Communication / Documentation — programme stored in case of power failure
Communication / Programming — AtmoCONTROL software for reading out, managing and organising the data logger via Ethernet interface (temporary trial version can be downloaded). USB stick with AtmoCONTROL software available as accessory (on demand).
Safety / Temperature control — adjustable electronic overtemperature monitor and mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20°C above nominal temperature
Safety / Autodiagnostic system — for fault analysis
Standard equipment / Works calibration certificate — Calibration at +160°C
Standard equipment / Door — fully insulated stainless steel doors with
2-point locking (compression door lock), lockable
Standard equipment / Internals — 2 stainless steel grid(s), electropolished
Stainless steel interior / Dimensions — w (A) x h (B) x d (C) : 1040 x 1200 x 600 mm (d less 39 mm for fan)
Stainless steel interior / Interior — easy-to-clean interior,made of stainless steel, reinforced by deep drawn ribbing with integrated and protected large-area heating on four sides
Stainless steel interior / Volume — 749 l
Stainless steel interior / Max. number of internals — 14
Stainless steel interior / Max. loading of chamber — 300 kg
Stainless steel interior / Max. loading per internal — 30 kg
Textured stainless steel casing / Dimensions — w (D) x h (E) x d (F) : 1224 x 1720 x 784 mm (d +56mm door handle)
Textured stainless steel casing / Installation — on lockable castors
Textured stainless steel casing / Housing — rear zinc-plated steel
Electrical data / Voltage
Electrical load — 400 V and 3x 230 V w/o neutral, 50/60 Hz approx. 7000 W