What is hot air sterilisation? Steriliser S makes use of dry sterilisation to ensure laboratory equipment and devices are sterilized. With this method, hot air, devoid of water vapor or moisture, makes sure the device is sterile without needing risk prone disinfection methods.
Temperature / Setting temperature range — +20 to +250 °C
Temperature / Setting accuracy temperature — up to 99.9 °C: 0.1 / from 100 °C: 0.5
Temperature / Working temperature range — at least 5 above ambient temperature to +250 °C
Temperature / Temperature sensor — 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error
Control technology / ControlCOCKPIT — TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
Control technology / Timer — Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
Control technology / Function HeatBALANCE — adapting the distribution of the heating performance of the upper and lower heating circuit from -50 % to +50 %
Control technology / Function SetpointWAIT — the process time does not start until the set temperature is reached
Control technology / Calibration — three freely selectable temperature values
Control technology / adjustable parameters — temperature (Celsius or Fahrenheit), air flap position, programme time, time zones, summertime/wintertime
Ventilation / Convection — natural convection
Ventilation / Fresh air — Admixture of pre-heated fresh air by electronically adjustable air flap
Ventilation / Vent — vent connection with restrictor flap
Communication / Documentation — programme stored in case of power failure
Communication / Programming — AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port
Safety / Temperature control — mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20°C above nominal temperature
Safety / Temperature control — overtemperature monitor TWW, protection class 3.1 or adjustable temperature limiter TWB, protection class 2, selectable on display
Safety / AutoSAFETY — additionally integrated over- and undertemperature monitor "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature
Safety / Autodiagnostic system — for fault analysis
Safety / Alarm — visual and acoustic
Standard equipment / Works calibration certificate — Calibration at +160°C
Standard equipment / Door — fully insulated stainless steel doors with
2-point locking (compression door lock), lockable
Standard equipment / Internals — 2 stainless steel grid(s), electropolished
Stainless steel interior / Interior — easy-to-clean interior,made of stainless steel, reinforced by deep drawn ribbing with integrated and protected large-area heating on four sides
Stainless steel interior / Volume — 449 l
Stainless steel interior / Dimensions — w (A) x h (B) x d (C) : 1040 x 720 x 600 mm
Stainless steel interior / Max. number of internals — 8
Stainless steel interior / Max. loading of chamber — 300 kg
Stainless steel interior / Max. loading per internal — 30 kg
Textured stainless steel casing / Dimensions — w (D) x h (E) x d (F) : 1224 x 1247 x 784 mm (d +56mm door handle)
Textured stainless steel casing / Installation — on lockable castors
Textured stainless steel casing / Housing — rear zinc-plated steel
Electrical data / Voltage
Electrical load — 400 V and 3x 230 V w/o neutral, 50/60 Hz approx. 5800 W