The 2300Na Sodium Analyzer provides early detection of trace sodium contamination in a water system with minimal operator intervention. It is an ideal solution for in-line sodium monitoring in ultrapure water or steam cycle chemistry.
Sodium Measurement Range — 0.001–100,000 ppb or equivalent ppm, autoranging
Sodium Measurement Accuracy — ±10% of reading or ±0.05 ppb, whichever is greater
Response Time — 90% at 5 min
Sample pH — 2.5–12
Update Rate — Once per second
Reagent Consumption — Diisopropylamine (DIPA), approximately 1 L per 3 months
Sample Flow Rate — >40 mL/min (>20 mL/min for cation exchange samples)
Sample Temperature — 5–50 °C (41–122 °F)
Sample Pressure — 0.3–7 bar (5–100 psig)
Grab Sample Measurement — Included
Power Supply — 100–240 VAC, 50–60 Hz, 25 W
Ambient Operating Temperature — 10–45 °C (50–113 °F)
Humidity — 10–90% non-condensing
Process Connection — 1/4 in or 6 mm OD tube stainless compression fitting; 19 x 25.4 mm drain hose, 2 m included
Operator Interface — 4-line backlit LCD, 5 tactile keys
Analog Output — Four powered 0/4–20 mA; 22 mA alarm; 500 ohm maximum load
Analog Output Scaling — Linear; bi-linear; logarithmic (1–4 decades); autoranging
Measurement Parameter — Sodium
Electrode Conditioning — Part of auto-calibration sequence
Sodium Calibration — Automatic unattended 3-point known addition
Calibration — Automatic
Analog Output Accuracy — ±0.05 mA
Model Family — 2300Na
Enclosure Type — Panel
Built-in Sequencer — No