Device type Circulation thermostat
Classification according to DIN 12876 III
Characteristics according to DIN 12877 FL
Refrigerant R290
Refrigerant quantity 79 g
Max. coolant pressure 21 bar
Heating power 1500 W
Cooling capacity (@20°C) 620 W
Cooling power (10 °C) 600 W
Cooling power (0 °C) 480 W
Cooling power (-10 °C) 350 W
Cooling power (-20 °C) 200 W
Cooling capacity (@ -30°C) 80 W
Operating temperature -30 - 100 °C
Min. operating temperature -30 °C
Temperature indicator yes
Temperature stability according to DIN 12876 (@ + 70 ° C) ±0.1 K
Temperature stability according to DIN 12876 (@ -10 °) ±0.1 K
PID temperature control
Working temperature sensor PT 100
Safety temperature sensor PT1000
Operating temperature indicator Diode line
Safe temperature indicator LED line
Temperature measurement, absolute accuracy Internal (adjustable by calibration) ±0.5 K
Accuracy of actual displayed temperature 0.1 K
Heating temperature fluctuation 0.1 K
Visual warning signal yes
Acoustic warning signal yes
Alarm for high temperature warning yes
Adjustable safe heating 0 - 110 °C
Automatic shutdown when the liquid level in the thermostat is below the permissible level yes
Overfill protection yes
bath volume 1.4 - 4.0 l
Pump type Pressure or exhaust pump
Pump capacity variable yes
Pump speed min. 2000 rpm
Pump speed max. 4000 rpm
Pump pressure max. (0 liter lifting water flow) 0.5 bar
Pressure pump (suction side) (flow 0 l) 0.25 bar
Max flow rate (0 bar counter pressure) 21 l/min
Fluid maximum viscosity 50 mm²/s
Pump connection M16x1
Possibility of calibration yes
Technical data according to DIN 12876
Allowed time on 100% condition
Noise level 58 dB(A)
Dimensions 220 x 480 x 525 mm
Weight 26 kg
Permissible ambient temperature 5 - 32 °C
Permissible relative humidity 80%
Protection class according to DIN EN 60529 IP 21
RS 232 connector yes
USB Micro-USB connector
Voltage 230 V
Frequency 50/60 Hz
Power consumption 1800 W