Temperature / Setting accuracy temperature — up to 99.9 °C: 0.1 / from 100 °C: 0.5
Temperature / Setting temperature range — +20 to +200 °C
Temperature / Temperature sensor — temperature measured through 4-wire Pt100 sensor individually for each thermoshelf
Temperature / Working temperature range — min. 5°C above ambient up to +200°C
Pressure (Vacuum) / Vacuum range — 5 to 1100 mbar
Pressure (Vacuum) / Pressure control — Digital electronic pressure control for a speed-controlled vacuum pump.
Tubing for vacuum, air and inert gas are made of material 1.4571 (ASTM 316 Ti). Programmable, digitally controlled inlet for air.
Pressure (Vacuum) / Permitted final vacuum — 0.01 mbar
Pressure (Vacuum) / Maximum leakage rate — 0.01 bar/h
Pressure (Vacuum) / Pump control — speed control, optimised rinsing procedures for the pump membranes as well as signal output for pump ON/OFF
Pressure (Vacuum) / Connection — Vacuum connection with small flange DN16, and gas inlet with fresh air supply
Control technology / adjustable parameters — temperature (Celsius or Fahrenheit), pressure (vacuum), programme time, time zones, summertime/wintertime
Control technology / Language setting — German, English, Spanish, French, Polish, Czech, Hungarian, Italian, Chinese
Control technology / ControlCOCKPIT — TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
Control technology / Timer — Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
Control technology / Function SetpointWAIT — the process time does not start until the set temperature is reached
Communication / Interface — Ethernet LAN,
USB
Communication / Documentation — programme stored in case of power failure
Communication / Programming — AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port
Safety / Temperature control — mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 10°C above nominal temperature
Safety / AutoSAFETY — additionally integrated over- and undertemperature protection "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature, compressor in case of undertemperature
Safety / Autodiagnostic system — integral fault diagnostics for temperature control
Safety / Alarm — visual and acoustic
Heating concept / VO direct heating — fuzzy-supported MLC (Multi-Level-Controlling) microprocessor controller adapting its performance to the volume (local temperature sensing) for each thermoshelf
Heating concept / Thermoshelves — 1 connection for thermoshelves in the rear
Standard equipment / Works calibration certificate — for +160°C at 20 mbar pressure for each supplied thermoshelf together with the vacuum oven
Standard equipment / Door — full-sight glass door, inside spring-loaded, 15 mm thick glazed panel in safety glass, outside with anti-splitter screen
Standard equipment / Internals — 1 thermoshelf of aluminim, material 3.3547 (ASTM B209) with integrated large-area heating
Stainless steel interior / Material — Hermetically welded stainless steel interior of extremely corrosion-resistant stainless steel, material 1.4404
Stainless steel interior / Interior — additional interior mountings of stainless steel, material 1.4404 (removable for cleaning), consisting of mounting at the sides with guide bars for thermoshelves and on top (diffusor) to avoid turbulences when aerating
Stainless steel interior / Volume — 29 l
Stainless steel interior / Dimensions — w (A) x h (B) x d (C) : 385 x 305 x 250 mm
Stainless steel interior / Max. number of internals — 2
Stainless steel interior / Max. loading of chamber — 40 kg
Stainless steel interior / Max. loading per internal — 20 kg
Textured stainless steel casing / Dimensions — w (D) x h (E) x d (F) : 550 x 607 x 400 mm (d +38mm door handle)
Textured stainless steel casing / Housing — rear zinc-plated steel
Electrical data / Voltage — Electrical load (maximally equipped) at 230 V, 50/60 Hz
Ambient conditions / Ambient temperature — +5 °C to +40 °C
Ambient conditions / Set Up — The distance between the wall and the rear of the chamber must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from the wall must not be less than 8 cm.
Ambient conditions / Humidity rh — max. 80 %, non-condensing
Ambient conditions / Overvoltage category — II
Ambient conditions / Pollution degree — 2
Packing/shipping data / Transport information — The appliances must be transported upright
Packing/shipping data / Customs tariff number — 8419 8998
Packing/shipping data / Country of origin — Federal Republic of Germany
Packing/shipping data / WEEE-Reg.-No. — DE 66812464
Packing/shipping data / Dimensions approx incl. carton — w x h x d: 660 x 870 x 590 mm
Packing/shipping data / Net weight — approx. 55 kg
Packing/shipping data / Gross weight carton — approx. 76 kg